:SBUS<n> Commands 24
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 539
:SBUS<n>:UART:TRIGger:DATA
(see page 798)
Command Syntax
:SBUS<n>:UART:TRIGger:DATA <value>
<value> ::= 8-bit integer from 0-255 (0x00-0xff) in decimal,
<hexadecimal>, <binary>, or <quoted_string> format
<hexadecimal> ::= #Hnn where n ::= {0,..,9 | A,..,F} for hexadecimal
<binary> ::= #Bnn...n where n ::= {0 | 1} for binary
<quoted_string> ::= any of the 128 valid 7-bit ASCII characters
(or standard abbreviations)
The :SBUS<n>:UART:TRIGger:DATA command selects the data byte value (0x00 to
0xFF) for the trigger QUALifier when in UART mode. The data value is used when
one of the RD or TD trigger types is selected.
When entering an ASCII character via the quoted string, it must be one of the 128
valid characters (case-sensitive): "NUL", "SOH", "STX", "ETX", "EOT", "ENQ",
"ACK", "BEL", "BS", "HT", "LF", "VT", "FF", "CR", "SO","SI", "DLE", "DC1", "DC2",
"DC3", "DC4", "NAK", "SYN", "ETB", "CAN", "EM", "SUB", "ESC", "FS","GS", "RS",
"US", "SP", "!", "\"", "#", "$", "%","&", "\'", "(", ")", "*", "+", ",", "-", ".", "/","0", "1", "2",
"3", "4", "5", "6", "7", "8", "9",":", ";", "<", "=", ">", "?", "@", "A", "B", "C","D", "E", "F",
"G", "H", "I", "J", "K", "L", "M","N", "O", "P", "Q", "R", "S", "T", "U", "V", "W", "X", "Y",
"Z","[", "\\", "]", "^", "_", "`", "a", "b", "c", "d", "e", "f", "g", "h", "i", "j", "k", "l", "m", "n",
"o", "p", "q", "r", "s", "t", "u", "v", "w", "x", "y", "z", "{", "|", "}", "~", or "DEL".
Query Syntax
:SBUS<n>:UART:TRIGger:DATA?
The :SBUS<n>:UART:TRIGger:DATA? query returns the current UART trigger data
value.
Return Format
<value><NL>
<value> ::= 8-bit integer in decimal from 0-255
See Also • "Introduction to :TRIGger Commands" on page 585
• ":TRIGger:MODE" on page 594
• ":SBUS<n>:UART:TRIGger:BASE" on page 537
• ":SBUS<n>:UART:TRIGger:TYPE" on page 542

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals