EasyManua.ls Logo

Keysight Technologies InfiniiVision DSOX1204G

Keysight Technologies InfiniiVision DSOX1204G
946 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
:TRIGger Commands 27
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 599
:TRIGger[:EDGE]:LEVel
(see page 798)
Command Syntax
:TRIGger[:EDGE]:LEVel <level>
<level> ::= <level>[,<source>]
<level> ::= 0.75 x full-scale voltage from center screen in NR3 format
for internal triggers
<level> ::= ±(external range setting) in NR3 format
for external triggers
<source> ::= {CHANnel<n> | EXTernal}
<n> ::= 1 to (# analog channels) in NR1 format
The :TRIGger[:EDGE]:LEVel command sets the trigger level voltage for the active
trigger source.
Query Syntax
:TRIGger[:EDGE]:LEVel? [<source>]
The :TRIGger[:EDGE]:LEVel? query returns the trigger level of the current trigger
source.
Return Format
<level><NL>
See Also "Introduction to :TRIGger Commands" on page 585
":TRIGger[:EDGE]:SOURce" on page 602
":EXTernal:RANGe" on page 257
NOTE
If the optional source is specified and is not the active source, the level on the active source is
not affected and the active source is not changed.