Common (*) Commands 5
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 119
*OPT (Option Identification)
(see page 798)
Query Syntax
*OPT?
The *OPT? query reports the options installed in the instrument. This query returns
a string that identifies the module and its software revision level.
Return Format
0,0,<license info>
<license info> ::= <All field>, <reserved>, <reserved>, <reserved>,
<Memory>, <Low Speed Serial>, <Automotive Serial>, <reserved>,
<reserved>, <reserved>, <Segmented Memory>, <Mask Test>, <reserved>,
<Bandwidth>, <reserved>, <reserved>, <reserved>, <reserved>,
<reserved>, <reserved>, <Educator's Kit>, <Waveform Generator>,
<reserved>, <reserved>, <reserved>, <reserved>, <reserved>,
<reserved>, <reserved>, <Digital Voltmeter>, <reserved>, <reserved>,
<reserved>, <Remote Command Logging>, <reserved>, <reserved>,
<reserved>
<All field> ::= {0 | All}
<reserved> ::= 0
<Memory> ::= {0 | MEMUP}
<Low Speed Serial> ::= {0 | EMBD}
<Automotive Serial> ::= {0 | AUTO}
<Segmented Memory> ::= {0 | SGM}
<Mask Test> ::= {0 | MASK}
<Bandwidth> ::= {0 | BW10 | BW20}
<Educator's Kit> ::= {0 | EDK}
<Waveform Generator> ::= {0 | WAVEGEN}
<Digital Voltmeter> ::= {0 | DVM}
<Remote Command Logging> ::= {0 | RML}
The *OPT? query returns the following:
See Also • "Introduction to Common (*) Commands" on page 109
• "*IDN (Identification Number)" on page 116
Module Module Id
No modules attached 0,0,0,0,0,0,0,0,0,0,0,0,0,SGM,MASK,0,0,0,0,0,0,0,0,EDK,WAVEGEN,0,0,0,
0,0,0,0,0,0,0,0,RML,0,0,0