EasyManua.ls Logo

Keysight Technologies InfiniiVision DSOX1204G

Keysight Technologies InfiniiVision DSOX1204G
946 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
:FUNCtion Commands 17
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 309
:FUNCtion:SOURce2
(see page 798)
Command Syntax
:FUNCtion:SOURce2 <value>
<value> ::= {CHANnel<n> | NONE}
<n> ::= 1 to (# analog channels) in NR1 format
The :FUNCtion:SOURce2 command specifies the second source for math
operations that have two sources (see the :FUNCtion:OPERation command), in
other words, ADD, SUBTract, or MULTiply. (The :FUNCtion:SOURce1 command
specifies the first source.)
If CHANnel1 or CHANnel2 is selected for :FUNCtion:SOURce1, the SOURce2
selection can be CHANnel1 or CHANnel2.
The :FUNCtion:SOURce2 setting is not used when the :FUNCtion:OPERation is FFT
(Fast Fourier Transform).
Query Syntax
:FUNCtion:SOURce2?
The :FUNCtion:SOURce2? query returns the currently specified second source for
math operations.
Return Format
<value><NL>
<value> ::= {CHAN<n> | NONE}
<n> ::= 1 to (# analog channels) in NR1 format
See Also "Introduction to :FUNCtion Commands" on page 291
":FUNCtion:OPERation" on page 303
":FUNCtion:SOURce1" on page 308

Table of Contents