:TRIGger Commands 27
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 605
:TRIGger:GLITch:LESSthan
(see page 798)
Command Syntax
:TRIGger:GLITch:LESSthan <less_than_time>[<suffix>]
<less_than_time> ::= floating-point number in NR3 format
<suffix> ::= {s | ms | us | ns | ps}
The :TRIGger:GLITch:LESSthan command sets the maximum pulse width duration
for the selected :TRIGger:GLITch:SOURce.
Query Syntax
:TRIGger:GLITch:LESSthan?
The :TRIGger:GLITch:LESSthan? query returns the pulse width duration time for
:TRIGger:GLITch:SOURce.
Return Format
<less_than_time><NL>
<less_than_time> ::= floating-point number in NR3 format.
See Also "Introduction to :TRIGger Commands" on page 585
":TRIGger:GLITch:SOURce" on page 610
":TRIGger:GLITch:QUALifier" on page 608
":TRIGger:MODE" on page 594