:SBUS<n> Commands 24
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 497
:SBUS<n>:LIN:SIGNal:BAUDrate
(see page 798)
Command Syntax
:SBUS<n>:LIN:SIGNal:BAUDrate <baudrate>
<baudrate> ::= integer from 2400 to 625000 in 100 b/s increments
The :SBUS<n>:LIN:SIGNal:BAUDrate command sets the standard baud rate of the
LIN signal from 2400 b/s to 625 kb/s in 100 b/s increments. If you enter a baud
rate that is not divisible by 100 b/s, the baud rate is set to the nearest baud rate
divisible by 100 b/s.
Query Syntax
:SBUS<n>:LIN:SIGNal:BAUDrate?
The :SBUS<n>:LIN:SIGNal:BAUDrate? query returns the current LIN baud rate
setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 2400 to 625000 in 100 b/s increments
See Also "Introduction to :TRIGger Commands" on page 585