Obsolete and Discontinued Commands 31
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 733
:MEASure:TDELta
(see page 798)
Query Syntax
:MEASure:TDELta?
The :MEASure:TDELta? query returns the time difference between the Tstop
marker (X2 cursor) and the Tstart marker (X1 cursor).
Tdelta = Tstop - Tstart
Tstart is the time at the start marker (X1 cursor) and Tstop is the time at the stop
marker (X2 cursor). No measurement is made when the :MEASure:TDELta? query
is received by the oscilloscope. The delta time value that is output is the current
value. This is the same value as the front-panel cursors delta X value.
Return Format
<value><NL>
<value> ::= time difference between start and stop markers in NR3 format
See Also • "Introduction to :MARKer Commands" on page 326
• "Introduction to :MEASure Commands" on page 350
• ":MARKer:X1Position" on page 329
• ":MARKer:X2Position" on page 331
• ":MARKer:XDELta" on page 333
• ":MEASure:TSTArt" on page 737
• ":MEASure:TSTOp" on page 738
NOTE
The :MEASure:TDELta command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:XDELta command (see page 333) instead.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals