204 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
10 :CHANnel<n> Commands
:CHANnel<n>:BANDwidth
(see page 798)
Command Syntax
:CHANnel<n>:BANDwidth <limit>
<limit> ::= 25E6 in NR3 format
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:BANDwidth command sets the bandwidth limit value and turns
on bandwidth limiting (see the :CHANnel<n>:BWLimit command).
For waveforms with frequencies below the bandwidth limit, turning the bandwidth
limit on removes unwanted high frequency noise from the waveform.
Bandwidth limit also limits the trigger signal path of the channel.
While you can request any limit; the oscilloscope will choose the only bandwidth
limit available, 25 MHz.
Query Syntax
:CHANnel<n>:BANDwidth? [MAXimum]
The :CHANnel<n>:BANDwidth? query returns the current setting of the low-pass
filter.
If the bandwidth limit is off, the query returns the full bandwidth of the
oscilloscope.
When the MAXimum parameter is used, the oscilloscope's maximum possible
bandwidth is returned.
Return Format
<limit><NL>
<limit> ::= 25E6 or full bandwidth in NR3 format
See Also • ":CHANnel<n>:BWLimit" on page 205

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals