:MEASure Commands 20
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 373
:MEASure:PPULses
(see page 798)
Command Syntax
:MEASure:PPULses [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:PPULses command installs a rising pulse count measurement on
screen. If the optional source parameter is not specified, the current source is
measured.
Query Syntax
:MEASure:PPULses? [<source>]
The :MEASure:PPULses? query measures and returns the on-screen rising pulse
count.
Return Format
<value><NL>
<value> ::= the rising pulse count in NR3 format
See Also "Introduction to :MEASure Commands" on page 350
":MEASure:SOURce" on page 380
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).