EasyManua.ls Logo

Keysight Technologies InfiniiVision DSOX1204G

Keysight Technologies InfiniiVision DSOX1204G
946 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
480 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
24 :SBUS<n> Commands
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth
(see page 798)
Command Syntax
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth <length>
<length> ::= integer from 1 to 8 in NR1 format
The :SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth command sets the number of
8-bit bytes in the CAN data string. The number of bytes in the string can be
anywhere from 1 bytes to 8 bytes (64 bits). The value for these bytes is set by the
:SBUS<n>:CAN:TRIGger:PATTern:DATA command.
Query Syntax
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth?
The :SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth? query returns the current
CAN data pattern length setting.
Return Format
<count><NL>
<count> ::= integer from 1 to 8 in NR1 format
Errors "-241, Hardware missing" on page 759
See Also "Introduction to :TRIGger Commands" on page 585
":SBUS<n>:CAN:TRIGger:PATTern:DATA" on page 479
":SBUS<n>:CAN:SOURce" on page 476

Table of Contents

Related product manuals