:CALibrate Commands 9
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 195
:CALibrate:OUTPut
(see page 798)
Command Syntax
:CALibrate:OUTPut <signal>
<signal> ::= {TRIGgers | MASK | OFF}
For oscilloscopes that have a built-in waveform generator (G-suffix models), the
:CALibrate:OUTPut command sets the signal that is available on the Gen Out BNC:
TRIGgers — pulse when a trigger event occurs.
MASK — signal from mask test indicating a failure.
OFF — either no signal or the waveform generator output signal when
:WGEN:OUTPut is ON.
Query Syntax
:CALibrate:OUTPut?
The :CALibrate:OUTPut query returns the Gen Out BNC signal selection.
Return Format
<signal><NL>
<signal> ::= {TRIG | MASK | OFF}
See Also "Introduction to :CALibrate Commands" on page 192
":WGEN:OUTPut" on page 691