306 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
17 :FUNCtion Commands
:FUNCtion:REFerence
(see page 798)
Command Syntax
:FUNCtion:REFerence <level>
<level> ::= the current reference level in NR3 format.
The :FUNCtion:REFerence command sets the voltage or vertical value represented
at center screen for the selected function. The range of legal values is generally
+/-10 times the current scale of the selected function, but will vary by function. If
you set the reference level to a value outside of the legal range, the level is
automatically set to the nearest legal value.
Query Syntax
:FUNCtion:REFerence?
The :FUNCtion:REFerence? query outputs the current reference level value for the
selected function.
Return Format
<level><NL>
<level> ::= the current reference level in NR3 format.
See Also • "Introduction to :FUNCtion Commands" on page 291
• ":FUNCtion:OFFSet" on page 302
• ":FUNCtion:RANGe" on page 305
• ":FUNCtion:SCALe" on page 307
NOTE
The FUNCtion:REFerence command is equivalent to the :FUNCtion:OFFSet command.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals