256 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
14 :EXTernal Trigger Commands
:EXTernal:PROBe
(see page 798)
Command Syntax
:EXTernal:PROBe <attenuation>
<attenuation> ::= probe attenuation ratio in NR3 format
The :EXTernal:PROBe command specifies the probe attenuation factor for the
external trigger. The probe attenuation factor may be 0.1 to 10000. This command
does not change the actual input sensitivity of the oscilloscope. It changes the
reference constants for scaling the display factors and for setting trigger levels.
If an AutoProbe probe is connected to the oscilloscope, the attenuation value
cannot be changed from the sensed value. Attempting to set the oscilloscope to
an attenuation value other than the sensed value produces an error.
Query Syntax
:EXTernal:PROBe?
The :EXTernal:PROBe? query returns the current probe attenuation factor for the
external trigger.
Return Format
<attenuation><NL>
<attenuation> ::= probe attenuation ratio in NR3 format
See Also • "Introduction to :EXTernal Trigger Commands" on page 249
• ":EXTernal:RANGe" on page 257
• "Introduction to :TRIGger Commands" on page 585
• ":CHANnel<n>:PROBe" on page 212

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals