:FRANalysis Commands 16
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 287
:FRANalysis:WGEN:VOLTage
(see page 798)
Command Syntax
:FRANalysis:WGEN:VOLTage <amplitude>
<amplitude> ::= amplitude in volts in NR3 format
The :FRANalysis:WGEN:VOLTage command specifies the waveform generator's
output sine wave amplitude.
Use the :WGEN:VOLTage:OFFSet command to specify the offset voltage or DC
level.
You can also specify the amplitude and offset indirectly using the
:WGEN:VOLTage:HIGH and :WGEN:VOLTage:LOW commands. For example, an
amplitude of 5 V and an offset of 1 V is the same as a high-level voltage of 4 V and
a low-level voltage of -1 V.
Query Syntax
:FRANalysis:WGEN:VOLTage?
The :FRANalysis:WGEN:VOLTage? query returns the currently specified waveform
generator amplitude.
Return Format
<amplitude><NL>
<amplitude> ::= amplitude in volts in NR3 format
See Also • ":FRANalysis:DATA" on page 275
• ":FRANalysis:ENABle" on page 276
• ":FRANalysis:FREQuency:STARt" on page 279
• ":FRANalysis:FREQuency:STOP" on page 280
• ":FRANalysis:SWEep:POINts" on page 284
• ":FRANalysis:RUN" on page 281
• ":FRANalysis:SOURce:INPut" on page 282
• ":FRANalysis:SOURce:OUTPut" on page 283
• ":FRANalysis:WGEN:LOAD" on page 286