:MARKer Commands 19
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 331
:MARKer:X2Position
(see page 798)
Command Syntax
:MARKer:X2Position <position> [suffix]
<position> ::= X2 cursor position in NR3 format
<suffix> ::= {s | ms | us | ns | ps | Hz | kHz | MHz}
The :MARKer:X2Position command:
• Sets :MARKer:MODE to MANual if it is not currently set to WAVeform (see
":MARKer:MODE" on page 328).
• Sets the X2 cursor position to the specified value.
X cursor units are set by the :MARKer:XUNits command.
Query Syntax
:MARKer:X2Position?
The :MARKer:X2Position? query returns current X2 cursor position. This is
functionally equivalent to the obsolete :MEASure:TSTOp command/query.
Return Format
<position><NL>
<position> ::= X2 cursor position in NR3 format
See Also • "Introduction to :MARKer Commands" on page 326
• ":MARKer:MODE" on page 328
• ":MARKer:X1Position" on page 329
• ":MARKer:X2Y2source" on page 332
• ":MARKer:XUNits" on page 334
• ":MEASure:TSTOp" on page 738
NOTE
If the front-panel cursors are off, the marker position values are not defined and an error is
generated. Make sure to set :MARKer:MODE to MANual or WAVeform to put the cursors in the
front-panel Normal mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals