:CHANnel<n> Commands 10
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 217
:CHANnel<n>:PROTection
(see page 798)
Command Syntax
:CHANnel<n>:PROTection[:CLEar]
<n> ::= 1 to (# analog channels) in NR1 format| 4}
With the InfiniiVision DSOX1204A/G oscilloscopes, the analog channel input
impedance is always 1 M
Ω, so automatic overvoltage protection is not necessary
(as it is for channels with 50
Ω input impedance). There are no protection settings
to clear, so the :CHANnel<n>:PROTection[:CLEar] command does nothing.
Query Syntax
:CHANnel<n>:PROTection?
The :CHANnel<n>:PROTection query always returns NORM (normal).
Return Format
NORM<NL>
See Also • "Introduction to :CHANnel<n> Commands" on page 203
• ":CHANnel<n>:COUPling" on page 206
• ":CHANnel<n>:PROBe" on page 212