EasyManua.ls Logo

Keysight Technologies InfiniiVision DSOX1204G

Keysight Technologies InfiniiVision DSOX1204G
946 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
:SBUS<n> Commands 24
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 473
:SBUS<n>:CAN:SAMPlepoint
(see page 798)
Command Syntax
:SBUS<n>:CAN:SAMPlepoint <value>
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
The :SBUS<n>:CAN:SAMPlepoint command sets the point during the bit time
where the bit level is sampled to determine whether the bit is dominant or
recessive. The sample point represents the percentage of time between the
beginning of the bit time to the end of the bit time.
Query Syntax
:SBUS<n>:CAN:SAMPlepoint?
The :SBUS<n>:CAN:SAMPlepoint? query returns the current CAN sample point
setting.
Return Format
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
See Also "Introduction to :TRIGger Commands" on page 585
":SBUS<n>:MODE" on page 465
":SBUS<n>:CAN:TRIGger" on page 477

Table of Contents

Related product manuals