:SBUS<n>:CAN:SAMPlepoint
(see page 798)
Command Syntax
:SBUS<n>:CAN:SAMPlepoint <value>
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
The :SBUS<n>:CAN:SAMPlepoint command sets the point during the bit time
where the bit level is sampled to determine whether the bit is dominant or
recessive. The sample point represents the percentage of time between the
beginning of the bit time to the end of the bit time.
Query Syntax
:SBUS<n>:CAN:SAMPlepoint?
The :SBUS<n>:CAN:SAMPlepoint? query returns the current CAN sample point
setting.
Return Format
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
See Also • "Introduction to :TRIGger Commands" on page 585
• ":SBUS<n>:MODE" on page 465
• ":SBUS<n>:CAN:TRIGger" on page 477