:TRIGger Commands 27
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 595
:TRIGger:NREJect
(see page 798)
Command Syntax
:TRIGger:NREJect <value>
<value> ::= {{0 | OFF} | {1 | ON}}
The :TRIGger:NREJect command turns the noise reject filter off and on. When the
noise reject filter is on, the trigger circuitry is less sensitive to noise but may
require a greater amplitude waveform to trigger the oscilloscope. This command is
not valid in TV trigger mode.
Query Syntax
:TRIGger:NREJect?
The :TRIGger:NREJect? query returns the current noise reject filter mode.
Return Format
<value><NL>
<value> ::= {0 | 1}