:MEASure Commands 20
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 389
:MEASure:VMAX
(see page 798)
Command Syntax
:MEASure:VMAX [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | FFT | WMEMory<r>}
<n> ::= 1 to (# of analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:VMAX command installs a screen measurement and starts a
maximum vertical value measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax
:MEASure:VMAX? [<source>]
The :MEASure:VMAX? query measures and outputs the maximum vertical value
present on the selected waveform.
Return Format
<value><NL>
<value> ::= maximum vertical value of the selected waveform in
NR3 format
See Also "Introduction to :MEASure Commands" on page 350
":MEASure:SOURce" on page 380
":MEASure:VMIN" on page 390
":MEASure:VPP" on page 391
":MEASure:VTOP" on page 394