EasyManuals Logo

Keysight Technologies InfiniiVision DSOX1204G User Manual

Keysight Technologies InfiniiVision DSOX1204G
946 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #334 background imageLoading...
Page #334 background image
334 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
19 :MARKer Commands
:MARKer:XUNits
(see page 798)
Command Syntax
:MARKer:XUNits <units>
<units> ::= {SEConds | HERTz | DEGRees | PERCent}
The :MARKer:XUNits command sets the X cursors units:
• SEConds — for making time measurements.
•HERTz — for making frequency measurements.
• DEGRees — for making phase measurements. Use the :MARKer:XUNits:USE
command to set the current X1 location as 0 degrees and the current X2
location as 360 degrees.
• PERCent — for making ratio measurements. Use the :MARKer:XUNits:USE
command to set the current X1 location as 0 percent and the current X2
location as 100 percent.
Changing X units affects the input and output values of the :MARKer:X1Position,
:MARKer:X2Position, and :MARKer:XDELta commands/queries.
Query Syntax
:MARKer:XUNits?
The :MARKer:XUNits? query returns the current X cursors units.
Return Format
<units><NL>
<units> ::= {SEC | HERT | DEGR | PERC}
See Also • "Introduction to :MARKer Commands" on page 326
• ":MARKer:XUNits:USE" on page 335
• ":MARKer:X1Y1source" on page 330
• ":MARKer:X2Y2source" on page 332
• ":MEASure:SOURce" on page 380
• ":MARKer:X1Position" on page 329
• ":MARKer:X2Position" on page 331

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals