334 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
19 :MARKer Commands
:MARKer:XUNits
(see page 798)
Command Syntax
:MARKer:XUNits <units>
<units> ::= {SEConds | HERTz | DEGRees | PERCent}
The :MARKer:XUNits command sets the X cursors units:
SEConds — for making time measurements.
•HERTz for making frequency measurements.
DEGRees — for making phase measurements. Use the :MARKer:XUNits:USE
command to set the current X1 location as 0 degrees and the current X2
location as 360 degrees.
PERCent — for making ratio measurements. Use the :MARKer:XUNits:USE
command to set the current X1 location as 0 percent and the current X2
location as 100 percent.
Changing X units affects the input and output values of the :MARKer:X1Position,
:MARKer:X2Position, and :MARKer:XDELta commands/queries.
Query Syntax
:MARKer:XUNits?
The :MARKer:XUNits? query returns the current X cursors units.
Return Format
<units><NL>
<units> ::= {SEC | HERT | DEGR | PERC}
See Also "Introduction to :MARKer Commands" on page 326
":MARKer:XUNits:USE" on page 335
":MARKer:X1Y1source" on page 330
":MARKer:X2Y2source" on page 332
":MEASure:SOURce" on page 380
":MARKer:X1Position" on page 329
":MARKer:X2Position" on page 331