EasyManua.ls Logo

Keysight Technologies InfiniiVision DSOX1204G

Keysight Technologies InfiniiVision DSOX1204G
946 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
734 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
31 Obsolete and Discontinued Commands
:MEASure:THResholds
(see page 798)
Command Syntax
:MEASure:THResholds {T1090 | T2080 | VOLTage}
The :MEASure:THResholds command selects the thresholds used when making
time measurements.
Query Syntax
:MEASure:THResholds?
The :MEASure:THResholds? query returns the current thresholds selected when
making time measurements.
Return Format
{T1090 | T2080 | VOLTage}<NL>
{T1090} uses the 10% and 90% levels of the selected waveform.
{T2080} uses the 20% and 80% levels of the selected waveform.
{VOLTage} uses the upper and lower voltage thresholds set by the
UPPer and LOWer commands on the selected waveform.
See Also "Introduction to :MEASure Commands" on page 350
":MEASure:LOWer" on page 731
":MEASure:UPPer" on page 740
NOTE
The :MEASure:THResholds command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MEASure:DEFine THResholds command (see page 356)
instead.

Table of Contents

Related product manuals