696 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
29 :WGEN Commands
:WGEN:VOLTage
(see page 798)
Command Syntax
:WGEN:VOLTage <amplitude>
<amplitude> ::= amplitude in volts in NR3 format
For all waveforms except DC, the :WGEN:VOLTage command specifies the
waveform's amplitude. Use the :WGEN:VOLTage:OFFSet command to specify the
offset voltage or DC level.
You can also specify the amplitude and offset indirectly using the
:WGEN:VOLTage:HIGH and :WGEN:VOLTage:LOW commands. For example, an
amplitude of 5 V and an offset of 1 V is the same as a high-level voltage of 4 V and
a low-level voltage of -1 V.
Query Syntax
:WGEN:VOLTage?
The :WGEN:VOLTage? query returns the currently specified waveform amplitude.
Return Format
<amplitude><NL>
<amplitude> ::= amplitude in volts in NR3 format
See Also "Introduction to :WGEN Commands" on page 671
":WGEN:FUNCtion" on page 673
":WGEN:VOLTage:OFFSet" on page 699
":WGEN:VOLTage:HIGH" on page 697
":WGEN:VOLTage:LOW" on page 698