:SBUS<n> Commands 24
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 505
:SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth
(see page 798)
Command Syntax
:SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth <length>
<length> ::= integer from 1 to 8 in NR1 format
The :SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth command sets the number of
8-bit bytes in the LIN data string. The number of bytes in the string can be
anywhere from 1 bytes to 8 bytes (64 bits). The value for these bytes is set by the
:SBUS<n>:LIN:TRIGger:PATTern:DATA command.
Query Syntax
:SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth?
The :SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth? query returns the current LIN
data pattern length setting.
Return Format
<count><NL>
<count> ::= integer from 1 to 8 in NR1 format
Errors "-241, Hardware missing" on page 759
See Also "Introduction to :TRIGger Commands" on page 585
":SBUS<n>:LIN:TRIGger:PATTern:DATA" on page 503
":SBUS<n>:LIN:SOURce" on page 498