EasyManuals Logo

Keysight Technologies InfiniiVision DSOX1204G User Manual

Keysight Technologies InfiniiVision DSOX1204G
946 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #505 background imageLoading...
Page #505 background image
:SBUS<n> Commands 24
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 505
:SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth
(see page 798)
Command Syntax
:SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth <length>
<length> ::= integer from 1 to 8 in NR1 format
The :SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth command sets the number of
8-bit bytes in the LIN data string. The number of bytes in the string can be
anywhere from 1 bytes to 8 bytes (64 bits). The value for these bytes is set by the
:SBUS<n>:LIN:TRIGger:PATTern:DATA command.
Query Syntax
:SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth?
The :SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth? query returns the current LIN
data pattern length setting.
Return Format
<count><NL>
<count> ::= integer from 1 to 8 in NR1 format
Errors • "-241, Hardware missing" on page 759
See Also • "Introduction to :TRIGger Commands" on page 585
• ":SBUS<n>:LIN:TRIGger:PATTern:DATA" on page 503
• ":SBUS<n>:LIN:SOURce" on page 498

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals