Obsolete and Discontinued Commands 31
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 723
:FRANalysis:PPDecade
(see page 798)
Command Syntax
:FRANalysis:PPDecade <value>
<value> ::= {10 | 20 | 30 | 40 | 50}
The :FRANalysis:PPDecade command specifies the number of points per decade in
the frequency response analysis.
Query Syntax
:FRANalysis:PPDecade?
The :FRANalysis:PPDecade? query returns the points per decade setting.
Return Format
<value><NL>
See Also • ":FRANalysis:SWEep:POINts" on page 284
• ":FRANalysis:FREQuency:STARt" on page 279
• ":FRANalysis:FREQuency:STOP" on page 280
This command is deprecated. You should now use the :FRANalysis:SWEep:POINts command
instead, which lets you specify the total number of points in the entire frequency sweep.
If you use this command, it will set the :FRANalysis:SWEep:POINts setting to a roughly
equivalent value.