Root (:) Commands 6
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 139
:AUToscale:AMODE
(see page 798)
Command Syntax
:AUToscale:AMODE <value>
<value> ::= {NORMal | CURRent}
The :AUTOscale:AMODE command specifies the acquisition mode that is set by
subsequent :AUToscales.
When NORMal is selected, an :AUToscale command sets the NORMal
acquisition type and the RTIMe (real-time) acquisition mode.
When CURRent is selected, the current acquisition type and mode are kept on
subsequent :AUToscales.
Use the :ACQuire:TYPE and :ACQuire:MODE commands to set the acquisition type
and mode.
Query Syntax
:AUToscale:AMODE?
The :AUToscale:AMODE? query returns the autoscale acquire mode setting.
Return Format
<value><NL>
<value> ::= {NORM | CURR}
See Also "Introduction to Root (:) Commands" on page 135
":AUToscale" on page 137
":AUToscale:CHANnels" on page 140
":ACQuire:TYPE" on page 189
":ACQuire:MODE" on page 181