EasyManua.ls Logo

Keysight Technologies InfiniiVision DSOX1204G

Keysight Technologies InfiniiVision DSOX1204G
946 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Obsolete and Discontinued Commands 31
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 717
:CHANnel2:SKEW
(see page 798)
Command Syntax
:CHANnel2:SKEW <skew value>
<skew value> ::= skew time in NR3 format
<skew value> ::= -100 ns to +100 ns
The :CHANnel2:SKEW command sets the skew between channels 1 and 2. The
maximum skew is +/-100 ns. You can use the oscilloscope's analog probe skew
control to remove cable delay errors between channel 1 and channel 2.
Query Syntax
:CHANnel2:SKEW?
The :CHANnel2:SKEW? query returns the current probe skew setting for the
selected channel.
Return Format
<skew value><NL>
<skew value> ::= skew value in NR3 format
See Also "Introduction to :CHANnel<n> Commands" on page 203
NOTE
The :CHANnel2:SKEW command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :CHANnel<n>:PROBe:SKEW command (see page 215)
instead.
NOTE
This command is only valid for the two channel oscilloscope models.

Table of Contents

Related product manuals