Obsolete and Discontinued Commands 31
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 717
:CHANnel2:SKEW
(see page 798)
Command Syntax
:CHANnel2:SKEW <skew value>
<skew value> ::= skew time in NR3 format
<skew value> ::= -100 ns to +100 ns
The :CHANnel2:SKEW command sets the skew between channels 1 and 2. The
maximum skew is +/-100 ns. You can use the oscilloscope's analog probe skew
control to remove cable delay errors between channel 1 and channel 2.
Query Syntax
:CHANnel2:SKEW?
The :CHANnel2:SKEW? query returns the current probe skew setting for the
selected channel.
Return Format
<skew value><NL>
<skew value> ::= skew value in NR3 format
See Also • "Introduction to :CHANnel<n> Commands" on page 203
The :CHANnel2:SKEW command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :CHANnel<n>:PROBe:SKEW command (see page 215)
instead.
This command is only valid for the two channel oscilloscope models.