180 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
8 :ACQuire Commands
:ACQuire:COUNt
(see page 798)
Command Syntax
:ACQuire:COUNt <count>
<count> ::= integer in NR1 format
In averaging mode, the :ACQuire:COUNt command specifies the number of values
to be averaged for each time bucket before the acquisition is considered to be
complete for that time bucket. When :ACQuire:TYPE is set to AVERage, the count
can be set to any value from 2 to 65536.
Query Syntax
:ACQuire:COUNT?
The :ACQuire:COUNT? query returns the currently selected count value for
averaging mode.
Return Format
<count_argument><NL>
<count_argument> ::= an integer from 2 to 65536 in NR1 format
See Also • "Introduction to :ACQuire Commands" on page 177
• ":ACQuire:TYPE" on page 189
• ":DIGitize" on page 143
• ":WAVeform:COUNt" on page 641
NOTE
The :ACQuire:COUNt 1 command has been deprecated. The AVERage acquisition type with a
count of 1 is functionally equivalent to the HRESolution acquisition type; however, you should
select the high-resolution acquisition mode with the :ACQuire:TYPE HRESolution command
instead.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals