:MEASure Commands 20
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 353
:MEASure:BRATe
(see page 798)
Command Syntax
:MEASure:BRATe [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r> | EXTernal}
<n> ::= 1 to (# of analog channels) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:BRATe command installs a screen measurement and starts the bit
rate measurement. If the optional source parameter is specified, the currently
specified source is modified.
Query Syntax
:MEASure:BRATe? [<source>]
The :MEASure:BRATe? query measures all positive and negative pulse widths on
the waveform, takes the minimum value found of either width type and inverts that
minimum width to give a value in Hertz.
Return Format
<value><NL>
<value> ::= the bit rate value in Hertz
See Also • "Introduction to :MEASure Commands" on page 350
• ":MEASure:SOURce" on page 380
• ":MEASure:FREQuency" on page 363
• ":MEASure:PERiod" on page 371
This command is not available if the source is FFT (Fast Fourier Transform).