738 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
31 Obsolete and Discontinued Commands
:MEASure:TSTOp
(see page 798)
Command Syntax
:MEASure:TSTOp <value> [suffix]
<value> ::= time at the stop marker in seconds
[suffix] ::= {s | ms | us | ns | ps}
The :MEASure:TSTOp command moves the stop marker (X2 cursor) to the
specified time with respect to the trigger time.
Query Syntax
:MEASure:TSTOp?
The :MEASure:TSTOp? query returns the time at the stop marker (X2 cursor).
Return Format
<value><NL>
<value> ::= time at the stop marker in NR3 format
See Also • "Introduction to :MARKer Commands" on page 326
• "Introduction to :MEASure Commands" on page 350
• ":MARKer:X1Position" on page 329
• ":MARKer:X2Position" on page 331
• ":MARKer:XDELta" on page 333
• ":MEASure:TDELta" on page 733
• ":MEASure:TSTArt" on page 737
NOTE
The short form of this command, TSTO, does not follow the defined Long Form to Short Form
Truncation Rules (see page 800). The normal short form "TST" would be the same for both
TSTArt and TSTOp, so sending TST for the TSTOp command produces an error.
NOTE
The :MEASure:TSTOp command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:X2Position command (see page 331) instead.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals