378 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
20 :MEASure Commands
:MEASure:SDEViation
(see page 798)
Command Syntax
:MEASure:SDEViation [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1-2 or 1-4 (# of analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:SDEViation command installs a screen measurement and starts std
deviation measurement. If the optional source parameter is specified, the current
source is modified.
Query Syntax
:MEASure:SDEViation? [<source>]
The :MEASure:SDEViation? query measures and outputs the std deviation of the
selected waveform. The oscilloscope computes the std deviation on all displayed
data points.
Return Format
<value><NL>
<value> ::= calculated std deviation value in NR3 format
See Also • "Introduction to :MEASure Commands" on page 350
• ":MEASure:VRMS" on page 392
• ":MEASure:SOURce" on page 380
NOTE
This ":MEASure:VRMS DISPlay, AC" command is the preferred syntax for making standard
deviation measurements.
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204G and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204G Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204G
CategoryTest Equipment
LanguageEnglish

Related product manuals