768 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
33 Status Reporting
The status register bits are described in more detail in the following tables:
• Table 43
• Table 41
• Table 48
• Table 49
• Table 51
• Table 46
The status registers picture above shows how the different status reporting data
structures work together. To make it possible for any of the Standard Event Status
Register bits to generate a summary bit, the bits must be enabled. These bits are
enabled by using the *ESE common command to set the corresponding bit in the
Standard Event Status Enable Register.
Wait
Trig
Run
Arm
Reg
AER ?
Ru n bi t se t if oscilloscope not stopped
:OPERation:CONDition?
Operation Status Condition Register
:OPEE
:OPEE?
Operation Status Enable (Mask) Register
10234567

OPER TRGMAVESB
RQS/
MSS
TRG
Reg
TER?
Trigger Event Register
*SRE
*SRE?
Service Request Enable (Mask) Register
*STB?
Status Byte Register
Outp ut
Queue
SRQ
*ESR?
(Standard) Event Status Register
*ESE
*ESE?
(Standard) Event Status Enable (Mask) Register
PON URQ EXE DDE QYE RQL OPCCM E
OVLR
USR
MSG

OR
10234567

OR
1023456798101112131415
Se r vi c e Re qu est
OR
Wait
Trig
Run
:OPERation[:EVENt]?
Operation Status Event Register
OVLR
11
5 3
From Overload
Event Registers
From Mask Test
Event Registers
MTE
MTE
9