EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Page 376

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysisMXO 4 Series
376User Manual 1335.5337.02 ─ 07
"Invert"
Inverts the current state of the frame and error types: all frames and
error types that were enabled are disabled and vice versa.
"Reset"
Resets the state of the selected frames and error types.
Remote command:
SBUS<sb>:CAN:FILTer:CHKall on page 777
SBUS<sb>:CAN:FILTer:CLR on page 777
SBUS<sb>:CAN:FILTer:INVert on page 778
SBUS<sb>:CAN:FILTer:RST on page 778
Frame type
Selects the frame type that you want to display. You can filter all enabled frame types
simultaneously.
For each frame type, you can also specify conditions for the value of the fields in the
"Edit" dialog.
Available frames are "CBFF", "CBFF-R", "CEFF", "CEFF-R", "FBFF", "FEFF", "XLFF",
"Overload" and "Error".
Remote command:
SBUS<sb>:CAN:FILTer:FRENable on page 780
SBUS<sb>:CAN:FILTer:FRAMe<fr>:ENABle on page 780
Edit
Opens a dialog to define the details of the selected frame.
CAN (option R&S MXO4-K520)

Table of Contents

Related product manuals