Remote control commandsMXO 4 Series
522User Manual 1335.5337.02 ─ 07
Return values:
<Attenuation> Range: 0.001 to 1000
Increment: 0.1
*RST: 1
Default unit: V/V
Usage: Query only
Asynchronous command
Manual operation: See "Attenuation" on page 127
PROBe<ch>:SETup:ATTenuation:MANual <Attenuation>
Sets the attenuation for an unknown probe.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<Attenuation> Range: 0.0001 to 10000
Increment: 0.1
*RST: 1
Default unit: V/V
Usage: Asynchronous command
Manual operation: See "Attenuation" on page 127
PROBe<ch>:SETup:ATTenuation:DEFProbe <PredefinedProbe>
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<PredefinedProbe> NONE | USER | ZC10 | ZC20 | ZC30 | ZD01A100 | ZD01A1000 |
ZS10L | ZC02100 | ZC021000 | ZC03
USER
Probe is not detected and not known to the instrument. Set unit
and attenuation manually.
ZC10 | ZC20 | ZC30 | ZS10L | ZC03
Type of the probe
ZD01A100 | ZD01A1000
High voltage differential probes, attenuation ratio according to
the setting on the probe.
A100 = 100:1
A1000 = 1000:1
ZC02100 | ZC021000
Current probes 100 A or 1000 A according to the setting on the
probe.
*RST:
NONE
Usage: Asynchronous command
Acquisition and setup