Remote control commandsMXO 4 Series
541User Manual 1335.5337.02 ─ 07
TRIGger:EVENt<ev>:TYPE <Type>
Selects the trigger type. In a trigger sequence the trigger type is set for each condition.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Type> EDGE | GLITch | WIDTh | RUNT | WINDow | TIMeout | INTerval |
SLEWrate | ANEDge | SETHold | STATe | PATTern
ANEDge = analog edge trigger is the only trigger type if the
extern trigger source is used.
For SETHold, also DATatoclock can be used.
*RST: EDGE
Usage: Asynchronous command
Manual operation: See "Type" on page 143
TRIGger:EVENt<ev>:LEVel<n>[:VALue]
<Level>
Sets the trigger level for the specified event and source (channel).
If the trigger source is serial bus, the trigger level is set by the thresholds in the proto-
col configuration.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
<n> 1 to 4, index of the analog channel
Parameters:
<Level> Range: -10 to 10
Increment: 0.001
*RST: 0
Default unit: V
Usage: Asynchronous command
Manual operation: See "[Level]" on page 40
TRIGger:FINDlevel
Sets the trigger level automatically to 0.5 * (MaxPeak – MinPeak).
In a trigger sequence, "Find level" affects all active events of the sequence (A, B, and
R event).
Usage:
Event
Asynchronous command
Manual operation: See "Find level" on page 144
Trigger