Remote control commandsMXO 4 Series
647User Manual 1335.5337.02 ─ 07
CALCulate:SPECtrum<sp>:FREQuency:BANDwidth[:RESolution][:VALue]
<RBW>
Queries or defines the used resolution bandwidth.
Suffix:
<sp>
.
1
Parameters:
<RBW> Range: 0.0002 to 2000000
Increment: 0.01
*RST: 2000000
Default unit: Hz
Usage: Asynchronous command
Manual operation: See "RBW" on page 249
CALCulate:SPECtrum<sp>:FREQuency:CENTer <Center>
Defines the position of the displayed frequency range, which is (Center - Span/2) to
(Center + Span/2). The width of the range is defined using the frequency span setting.
Suffix:
<sp>
.
1
Parameters:
<Center> Range: -1E+26 to 1E+26
Increment: 0.01
*RST: 0
Default unit: HZ
Usage: Asynchronous command
Manual operation: See "[Position]" on page 41
CALCulate:SPECtrum<sp>:FREQuency:SCALe <XAxisMode>
Defines the scaling method for the frequency axis (x-axis) of the spectrogram.
Suffix:
<sp>
.
1
Parameters:
<XAxisMode> LIN | LOG
LIN: linear scaling
LOG: logarithmic scaling
*RST: LIN
Usage: Asynchronous command
Manual operation: See "Frequency axis" on page 251
Spectrum analysis