Remote control commandsMXO 4 Series
740User Manual 1335.5337.02 ─ 07
Return values:
<AddrComplete> ON | OFF
*RST: OFF
Usage: Query only
Asynchronous command
SBUS<sb>:I2C:FRAMe<fr>:ADBStart?
Returns the start time of the address acknowledge bit.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<AddrAckBtStrt> Range: -1E+26 to 1E+26
Increment: 1E-10
*RST: 0
Default unit: s
Usage: Query only
Asynchronous command
SBUS<sb>:I2C:FRAMe<fr>:ADDRess?
Returns the device address value of the indicated frame. That is, the address value
that is shown in the decoded cells and in the decode results table.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<AddressValue> Range: 0 to 2047
Increment: 1
*RST: 0
Usage: Query only
Asynchronous command
SBUS<sb>:I2C:FRAMe<fr>:ADEVice?
Returns the pure device address of the indicated frame without the R/W bit.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Protocols