Remote control commandsMXO 4 Series
834User Manual 1335.5337.02 ─ 07
<fr> index of the frame
<fl> index of the field
Parameters:
<Index_Max> Range: 1 to 65535
Increment: 1
*RST: 65535
Usage: Asynchronous command
Manual operation: See "Edit" on page 417
SBUS<sb>:TNOS:FILTer:IMIN
<Frame>,<Field>,<Data>
SBUS<sb>:TNOS:FILTer:FRAMe<fr>:FLD<fl>:IMIN <Index_Min>
Specifies the index, or sets the start value of an index range.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
<fl> index of the field
Parameters:
<Index_Min> Range: 1 to 65535
Increment: 1
*RST: 1
Usage: Asynchronous command
Manual operation: See "Edit" on page 417
SBUS<sb>:TNOS:FILTer:IOPerator <Frame>,<Field>,<Operator>
SBUS<sb>:TNOS:FILTer:FRAMe<fr>:FLD<fl>:IOPerator <Index_Operator>
Sets the operator for the index in the selected field of the selected frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
<fl> index of the field
Parameters:
<Index_Operator> EQUal | INRange | RANGe
*RST: INRange
Usage: Asynchronous command
Manual operation: See "Edit" on page 417
Protocols