Remote control commandsMXO 4 Series
845User Manual 1335.5337.02 ─ 07
Parameters:
<Skew>
Usage: Asynchronous command
DIGital<m>:STATe <State>
Enables or disables the indicated digital channel, displays it, and enables the Logic 1 if
the bus was disabled.
If another active bus already uses the selected digital channel, the instrument disables
the other bus to avoid conflicts.
For Logic 1, the DIG::STAT command has the same effect as PBUS<pb>:STATe. To
enable digital channels for buses 2, 3 and 4, use the PBUS:BIT[:STAT] command.
Suffix:
<m>
.
0..15
Selects the digital channel.
Parameters:
<State> ON | OFF
Usage: Asynchronous command
DIGital<m>:TECHnology <Technology>
Selects the threshold voltage for various types of integrated circuits and applies it to all
digital channels.
The setting affects only the settings of the first MSO bus (Logic1).
You can set the technology value for all buses with PBUS<pb>:TECHnology.
Suffix:
<m>
.
0..15
The suffix is irrelevant.
Parameters:
<Technology> V15 | V25 | V165 | V125 | V09 | VM13 | V38 | V20 | V0 | MANual
See PBUS<pb>:TECHnology.
Usage: Asynchronous command
DIGital<m>:THCoupling <State>
Sets the threshold and the hysteresis for all digital channels of Logic1 to the same
value.
Suffix:
<m>
.
0..15
The suffix is irrelevant.
Mixed signal option (option R&S MXO4-B1)