Power analysis (option R&S RTE-K31)
R&S
®
RTE
1019User Manual 1326.1032.02 ─ 20
The measured peaks have different origin. Analyzing the frequencies gives information
about the influences on the output signal.
The following remote commands are used for handling the measurement results:
●
POWer:SPECtrum:RCOunt? on page 1982
●
POWer:SPECtrum:RESult<m>:FREQuency? on page 1982
●
POWer:SPECtrum:RESult<m>:LEVel? on page 1982
●
POWer:SPECtrum:REPort:ADD on page 1982
16.6.3.2 Configuring output spectrum
For details of the configuration settings, see Chapter 16.6.3.3, "Output spectrum set-
tings", on page 1020.
1. Select "Analysis">"Power".
2. Under "Output", select "Spectrum".
3. Select the "Channels" tab.
4. Connect the probe to the DUT and to the oscilloscope as shown in the graphic:
5. Select the correct channel for the "Voltage Source".
6. Select the "Settings" tab.
7. Set the "SMPS switching frequency" according to your signal.
8. Select an "Optimize Scaling"option.
9. Tap "Execute".
10. Set the positions of the cursors according to the measured spectrum.
Output