Remote control commands
R&S
®
RTE
1269User Manual 1326.1032.02 ─ 20
<n> 1..30
Number of the eye/jitter measurement, see Table 18-6.
Parameters:
<UpperLimit> Range: -100 to 100
Increment: 10E-6
*RST: 0
Usage: Asynchronous command
MEASurement<m>:EYEJitter:LCHeck<n>:LOWer:MARGin <LowerMargin>
MEASurement<m>:EYEJitter:LCHeck<n>:UPPer:MARGin <UpperMargin>
Defines the upper margin for the limit check. Margins are not as strict as limits and
must be within the valid value range. The valid range is defined using the
MEASurement<m>:EYEJitter:LCHeck<n>:VALid command.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
<n> 1..30
Number of the eye/jitter measurement, see Table 18-6.
Parameters:
<UpperMargin> Range: -100 to 100
Increment: 10E-6
*RST: 0
Usage: Asynchronous command
18.12.5 Spectrum
The following table lists the <MeasType> parameter values with a short description.
In addition, the suffixes <n> for limit check remote commands are listed.
For a detailed description, see Chapter 8.2.7.1, "Overview of spectrum measure-
ments", on page 364.
Table 18-7: Spectrum measurements
<MeasType>
parameter value
(remote control)
Measurement
(manual opera-
tion)
Description/Result Suffix
<n>
CPOWer Channel power Power integrated over the sample values defined by
a center frequency and a bandwidth; based on a
defined impedance; the result is given in mW
1
not used 2
OBWidth Occupied band-
width
From the defined center frequency, symmetric sam-
ple value pairs to the left and right are integrated
until a user-defined percentage of the total power is
reached
3
Automatic measurements