EasyManua.ls Logo

Rohde & Schwarz RTE Series - Time Scale;Acquisition Time; Acquisition Time; Time Scale

Rohde & Schwarz RTE Series
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Acquisition and waveform setup
R&S
®
RTE
177User Manual 1326.1032.02 ─ 20
Time scale...................................................................................................................177
Acquisition time...........................................................................................................177
Reference point...........................................................................................................178
Position....................................................................................................................... 178
Restrict horizontal position to acquisition range..........................................................178
Sample rate.................................................................................................................178
Resolution................................................................................................................... 178
Record length..............................................................................................................178
ZVC Resolution...........................................................................................................179
ZVC Record length......................................................................................................179
Resolution / Record length (Resolution dependency).................................................179
Auto adjustment (Resolution dependency)................................................................. 179
Record length limit (Resolution dependency)............................................................. 179
Roll mode....................................................................................................................180
Minimum acquisition time..............................................................................181
Time scale
Sets the horizontal scale for all channel and math waveforms in seconds per division.
Increase the scale to see a longer time interval of the waveform. Decrease the scale to
see it in more detail. The scale has a point that remains fixed on the screen when the
scale value is changing - the reference point.
Remote command:
TIMebase:SCALe on page 1108
Acquisition time
Shows the time of one acquisition, that is the time across the 10 divisions of the dia-
gram:
Acquisition time = Time scale * 10 divisions
Changing the acquisition time changes the time scale too.
Remote command:
TIMebase:RANGe on page 1108
Horizontal settings

Table of Contents

Other manuals for Rohde & Schwarz RTE Series

Related product manuals