Protocol analysis
R&S
®
RTE
697User Manual 1326.1032.02 ─ 20
3. In the "Decode" tab, select the protocol: "Ethernet".
4. Tap "Ethernet Variant" and select the variant ("10BASE-T" or "100BASE-TX") you
want to set up.
Note: Note that no triggering on the serial bus is available.
To trigger the signal, use the edge trigger on the source channel.
5. For the variant "10BASE-T", select the polarity ("Normal" or "Inverted") of the data
signal.
6. Set the logical thresholds.
7. Enable "Decode".
13.11.3 Ethernet trigger
If you need information on how to get started with triggering on Ethernet 10BASE-T
signals, see Chapter 13.11.3.2, "Triggering on Ethernet signals", on page 702. Other-
wise proceed with the Ethernet trigger settings.
13.11.3.1 Ethernet trigger settings
Access: [PROTOCOL] > "Trigger" tab > "Source = Serial Bus"
Ethernet 10BASE-T and 100BASE-TX (option R&S RTE-K8)