Remote control commands
R&S
®
RTE
1295User Manual 1326.1032.02 ─ 20
MEASurement<m>:LTMeas:TIME <MeasurementTime>
Defines the total duration of the long term measurement.
This setting is only available if MEASurement<m>:STATistics:MODE is set to
"Time".
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<MeasurementTime> Range: 0.01 to 2.14748E+9
Increment: 1
*RST: 200
Default unit: s
Usage: Asynchronous command
18.12.9
Protocol
Before starting the protocol measurement, a bus type has to be selected (BUS<m>:
TYPE) and the bus has to be activated (BUS<m>[:STATe]).
The following table lists the <MeasType> parameter values with a short description.
Table 18-9: Protocol measurement types
<MeasType> param-
eter value (remote
control)
Measurement (man-
ual operation)
Description/Result
AUDio Audio
FLDValue Field value Value of a field over time
F2F Frame to frame The distance between two frame types
Trigger to frame
GAP Gap Measures a gap, periods at which the bus is idle. The dis-
tance of a gap can only be measured between two identi-
fied frames.
MBITrate Main bit rate Bit rate of the protocol
SBITrate 2nd bit rate Additional bit rate, protocol dependent.
BIDLe Bus idle Calculates the bus idle time
FCNT Frame count Number of all frames within the acquisition window
FEC Frame error count Sum of all frames with errors within the acquisition window