Remote control commands
R&S
®
RTE
1967User Manual 1326.1032.02 ─ 20
POWer:ONOFf:RESult:TOFF?
POWer:ONOFf:RESult:TON?
Returns the result time.
Return values:
<TurnOnTime> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
POWer:ONOFf:TIME
<Time>
Sets the time, the start of the measurement of the turn off time is delay with, after the
trigger point.
Parameters:
<Time> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 0.1
Default unit: s
POWer:ONOFf:TYPE <MeasType>
Selects the measurement type.
Parameters:
<MeasType> TON | TOFF
TON - "Turn on" measurement
TOFF - "Turn off" measurement
*RST: TON
18.20.12 Switching loss
POWer:SWITching:AUTO............................................................................................. 1968
POWer:SWITching:EXECute......................................................................................... 1968
POWer:SWITching:REPort:ADD.................................................................................... 1968
POWer:SWITching:SWIFrequency.................................................................................1968
POWer:SWITching:SWIT.............................................................................................. 1969
POWer:SWITching:COND.............................................................................................1969
POWer:SWITching:NCON.............................................................................................1969
POWer:SWITching:TON................................................................................................1969
POWer:SWITching:TOFF..............................................................................................1969
POWer:SWITching:TOTal..............................................................................................1969
POWer:SWITching:GATE:COND:STARt......................................................................... 1969
POWer:SWITching:GATE:COND:STOP..........................................................................1969
POWer:SWITching:GATE:NCON:STARt......................................................................... 1969
POWer:SWITching:GATE:TOFF:STARt.......................................................................... 1969
Power analysis (option R&S
RTE-K31)