EasyManua.ls Logo

Rohde & Schwarz RTE Series - Probe Setup: Extern Tab

Rohde & Schwarz RTE Series
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Triggers
R&S
®
RTE
274User Manual 1326.1032.02 ─ 20
2. Set up the probe: "Trigger" menu > "Extern" tab.
See Chapter 6.8.1, "Probe setup: Extern tab", on page 274.
3. Set up the trigger:
a) Select the "Trigger" > "Setup" tab.
b) Select the source: "Extern"
c) Adjust the trigger settings.
See Chapter 6.8.2, "External trigger setup", on page 275.
If the trigger source is a channel input, the trigger system uses the digitized signal. The
trigger system of the instrument is a separate system, thus the signal processing by
enhancement, decimation and arithmetic has no impact on the trigger signal. Most of
the R&S RTE trigger types use the digitized trigger signal.
If the trigger source is the external trigger input, the trigger comparator uses the analog
input signal. For the external trigger signal, only the edge trigger of the A-trigger is
available. Trigger sequence is not supported.
6.8.1 Probe setup: Extern tab
The "Trigger" > "Extern" tab provides all settings that are relevant for the probe that is
connected to the external trigger input. The functionality on the tab changes according
to the type of the attached probe. They are the same as for probe setup of input chan-
nels: see Chapter 5.5, "Probes", on page 193.
If the external trigger input is overloaded, a message informs you.
Remote commands:
TRIGger<m>:EXTern:OVERload on page 1210
TRPRobe:ID:PARTnumber? on page 1141
TRPRobe:SETup:STATe? on page 1124
TRPRobe:SETup:TYPE? on page 1124
TRPRobe:SETup:ATTenuation:DEFProbe on page 1138
External trigger input

Table of Contents

Other manuals for Rohde & Schwarz RTE Series

Related product manuals