Triggers
R&S
®
RTE
248User Manual 1326.1032.02 ─ 20
Upper level
Sets the upper voltage threshold.
Remote command:
TRIGger<m>:LEVel<n>:RUNT:UPPer on page 1174
Lower level
Sets the lower voltage threshold.
Remote command:
TRIGger<m>:LEVel<n>:RUNT:LOWer on page 1174
Range
Selects how the time limit of the runt pulse is defined:
"Any runt"
Triggers on all runts fulfilling the level condition, without time limita-
tion.
"Longer"
Triggers on runts longer than the given "Runt width".
"Shorter"
Triggers on runts shorter than the given "Runt width".
"Within"
Triggers if the runt length is inside a given time range. The range is
defined by "Runt width" and "±Delta".
"Outside"
Triggers if the runt length is outside a given time range. The range
definition is the same as for "Within" range.
Remote command:
TRIGger<m>:RUNT:RANGe on page 1174
Runt width
For the ranges "Shorter" and "Longer", the runt width defines the maximum and mini-
mum pulse width, respectively.
For the ranges "Within" and "Outside", the runt width defines the center of a range
which is defined by "±Delta".
Remote command:
TRIGger<m>:RUNT:WIDTh on page 1175
±Delta
Defines a range around the given runt width.
Remote command:
TRIGger<m>:RUNT:DELTa on page 1175
6.3.6 Window
Access: [TRIGGER] > "Setup" tab > "Type = Window"
The window trigger checks the signal run in relation to a "window". The window is
formed by the upper and lower voltage levels. The trigger condition is fulfilled, if the
waveform enters or leaves the window, or if the waveform stays inside or outside for a
time longer or shorter than specified.
With the window trigger, you can display longer transient effects.
Trigger types