● Setup parameters of modular probes....................................................................207
● Termination voltage with R&S RT-ZMA40 SMA module....................................... 208
5.5.5.1 Setup parameters of modular probes
The basic setup parameters of all modular probes are the measurement mode and the
offset settings.
Access: "Vertical" menu > "Probe Setup"
Probe mode
Sets the measurement mode of modular probes.
The modular probes of the R&S RT-ZM family have a multi-mode function. You can
switch between single-ended, differential and common mode measurements without
reconnecting or resoldering the probe. You can set the probe mode in the dialog box,
and you can assign the probe mode setting to the micro button.
If you use the R&S RT-ZMA30 browser module, only DM measurements are possible
because this module has no ground connector.
The measurement modes are:
"DM"
Differential mode input voltage (V
dm
), the voltage between the positive
and negative input terminal.
"CM"
Common mode input voltage (V
cm
), the mean voltage between the
positive and negative input terminal vs. ground.
Probes