Mask testing
R&S
®
RTE
432User Manual 1326.1032.02 ─ 20
Vertical stretch
Sets the vertical scaling to stretch the mask in y-direction. The scaling axis is the hori-
zontal line through the lowest value of the lower mask limit. Values > 100% stretch the
mask, and values < 100% compress it.
Figure 10-4: Waveform mask with vertical width = 0.5 div, vertical position = -0.5 div, vertical stretch =
110%
Remote command:
MTESt:WFMRescale:YSTRetch on page 1343
Vertical position
Moves the mask vertically within the display.
Figure 10-5: Waveform mask with vertical width = 0.5 div and vertical position = -0.5 div
Remote command:
MTESt:WFMRescale:YPOSition on page 1343
Reference waveform: save to or load from file
Loads the waveform from the selected file to the "Reference" and creates the mask
immediately.
See also: "Save to or load from file" on page 293.