Remote control commands
R&S
®
RTE
1160User Manual 1326.1032.02 ─ 20
Vertical settings of P and N output are the vertical settings of the input channels.
Suffix:
<m>
.
1..2
Differential signal
Parameters:
<Position> Position value
Range: -5 to 5
Increment: 0.02
*RST: 0
Default unit: div
Usage: Asynchronous command
18.8.10
Digital filter
CHANnel<m>:DIGFilter:STATe.......................................................................................1160
TRIGger<m>:RFReject................................................................................................. 1160
TRIGger<m>:RFSReject...............................................................................................1161
CHANnel<m>:DIGFilter:STATe <State>
Enables the DSP filter for input channels.
Suffix:
<m>
.
1..4
Irrelevant, omit the suffix. The filter is enabled to all channels.
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
TRIGger<m>:RFReject <Bandwidth>
Sets the limit frequency for input channels and the trigger source.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<Bandwidth> Range: 100E+3 to 2E+9
Increment: 1000
*RST: 1E+6
Default unit: Hz
Usage: Asynchronous command