Remote control commands
R&S
®
RTE
1159User Manual 1326.1032.02 ─ 20
Parameters:
<Scale> Scale value, given in Volts per division.
Range: 1E-3 to 100
Increment: 1E-3
*RST: 0.05
Default unit: V/div
Usage: Asynchronous command
DIFFerential<m>:COUPling <ScaleCoupling>
If enabled for a differential signal, the vertical scales of P, N, differential and common
mode waveforms are coupled. The scales are related as follows:
VertScale
P
= VertScale
N
= VertScale
CM
= VertScale
Diff
/2
You can disable the scale coupling and set an individual scale for each waveform.
Suffix:
<m>
.
1..2
Differential signal
Parameters:
<ScaleCoupling> ON | OFF
*RST: ON
Usage: Asynchronous command
DIFFerential<m>:COMMon:OFFSet <Offset>
DIFFerential<m>:DIFFerential:OFFSet <Offset>
Set the offset of differential and common mode waveforms.
Vertical settings of P and N output are the vertical settings of the input channels.
Suffix:
<m>
.
1..2
Differential signal
Parameters:
<Offset> Offset value
Range: -1 to 1
Increment: 0.01
*RST: 0
Default unit: V
Usage: Asynchronous command
DIFFerential<m>:COMMon:POSition <Position>
DIFFerential<m>:DIFFerential:POSition <Position>
Set the vertical position of differential and common mode waveforms.
Acquisition and setup